Sign In | Join Free | My fazendomedia.com |
|
Brand Name : Yokogawa
Model Number : FLXA21-D-P-D-AB-C1-NN-A-N-LA-N-NN
Certification : -
Place of Origin : -
MOQ : 1
Price : USD650/PC
Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability : 1000
Delivery Time : 6-12WEEKS
Packaging Details : Original
Brand : Yokogawa
Type : FLXA21-D-P-D-AB-C1-NN-A-N-LA-N-NN
Color : white
Ambient Temperature : -20 to 55°C
Control Drawing : Refer to App. 5.1
Home display (when two sensors are connected) 6.35 Tag:PH 24.9 4mA PH1 20mA °C 24 mV pH Tag:PH 25.0 °C 19 mV pH 10.38 A B C F K H E D G J Figure 1.8 Example of home display Home display The Home display appears upon startup when two sensors are connected and the MONITOR display is disabled. (Home display is not available when only one sensor is connected) A: Measurement value: Primary value (large characters/user selectable) B, C: Measurement value: Secondary and tertiary values (small characters) D: Unit for the primary value E: Tag No. (user programmable) F: Sensor No. G: Sensor wellness indicator (More ■ indicate the better condition.) H: Hold/Wash indicators (appear only during the Hold/Wash operations) J: Analog output display and parameter (ex.: PH1…PH=Parameter, 1=sensor number) K: Fault/Warning indicators (indicated in blinking only during Fault/Warning status) 4mA Diff-pH 20mA pH 0.03 C 6.35 24.9 °C 24 mV Tag:PH pH Calculated data 4mA PH1 20mA pH 10.38 R(1) 25.0 °C 19 mV pH 10.38 Tag:PH Redundant system Figure 1.9 Example of home display of calculated data and redundant system n Zoom display Next 20 12 4 mA 15.00 HOLD FAIL WASH Next Figure 1.10 Example zoom display The zoom display appears when the Zoom button on the main display is pressed. The Zoom display shows an easy-to-read graphical display of the output status. Pressing Next displays the contact state if communication is set to PH201G distributor. Pressing Next again displays the status information of the instrument. <1. INTRODUCTION AND GENERAL DESCRIPTION> 1-9 IM 12A01A02-01E 9th Edition : Mar. 23, 2018-00 n Trend display 12.00 8.40 5.60 2.50 12:00 12:20 12:40 9.10pH Minimum Maximum Average Current process value Time Tag: Measurement value (pH) Maximum value on this display Minimum value on this display PH pH1 A D B G E C F Figure 1.11 Example of trend display Trend display The Trend display appears when the Trend button on the Zoom display is pressed. A: X axis; Time axis (user programmable from 15 minutes to 14 days) B: Y axis; Measurement value axis (user programmable) C: Tag No. D: Current measurement value with unit E: Trend (maximum, minimum and average values during the display update time) F: Icon (current measurement value, and maximum and minimum values until the display update) G: Fault/Warning indicators (indicated in blinking only during Fault/Warning status
Marking/Rating | Nonincendive for Class I |
Ambient Humidity | 0 – 100% |
Enclosure | IP66 |
![]() |
Original Yokogawa FLEXA Series Analyzers FLXA21-D-P-D-AB-C1-NN-A-N-LA-N-NN Low Price Images |